FDA Adverse Event Injury Summary report: N

THREADED GUIDE WIRE W/ TROCAR TIP, DIA 1.35MM

MDR report key: 2190532 · Received August 4, 2011

Report

Report Number
1220246-2011-00118
Event Type
Injury
Date Received
August 4, 2011
Date of Event
July 5, 2011
Report Date
July 8, 2011
Manufacturer
ARTHREX, INC.
Product Code
MDM
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

PATIENT DEMOGRAPHICS (AGE AT TIME OF EVENT, DATE OF BIRTH, GENDER, WEIGHT) WERE REQUESTED BUT NOT PROVIDED. NO FURTHER PATIENT INFORMATION WAS PROVIDED AT THE TIME OF THIS REPORT OR MADE AVAILABLE IN RESPONSE TO FOLLOW-UP COMMUNICATION. NO ADDITIONAL ADVERSE CONSEQUENCES HAVE BEEN REPORTED FROM THIS EVENT. THIS DEVICE IS USED FOR TREATMENT. THE DEVICE WAS REQUESTED FOR EVALUATION BUT PART REMAINS IN THE PATIENT AND CANNOT BE RETURNED THEREFORE THE COMPLAINANT'S EVENT COULD NOT BE VERIFIED. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION. THE MOSTLY LIKELY CAUSES OF THE K-WIRE BREAKING IS DUE TO NOT FOLLOWING THE PROPER TECHNIQUE AND PRYING/LEVERAGING OF THE K-WIRE. LOT NUMBER WAS REQUESTED BUT NOT PROVIDED; THEREFORE DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED... THE POTENTIAL CAUSES OF THIS EVENT ARE BEING COMMUNICATED TO THE EVENT REPORTER. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE THREADED TIPS OF THE 2 K-WIRES BROKE ON INSERTION DURING ANKLE FRACTURE REPAIR,(TIPS NOT RETRIEVED). A 2.6 DRILL WAS USED FOR TUNNEL PREP PRIOR TO INSERTION. THREADED TIPS ARE DEEP IN THE BONE AND WOULD NOT RELEASE AS SURGEON TRIED TO REMOVE THE WIRES, BOTH OF THE IMPLANTS WERE LEFT IN PLACE AND ARE OVERTOP OF THE BROKEN TIP PIN. SURGEON USED A SIDE TO SIDE MOTION WHILE TRYING TO REMOVE THE WIRES. BONE QUALITY HARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THREADED GUIDE WIRE W/ TROCAR TIP, DIA 1.35MM INSTRUMENT, MANUAL, SURGICAL, GENERAL USE MDM ARTHREX, INC. UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Other