FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 21905278 · Received April 24, 2025

Report

Report Number
9617229-2025-06816
Event Type
Injury
Date Received
April 24, 2025
Date of Event
January 1, 2017
Report Date
June 20, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS IS A FOLLOW-UP REPORT TO A MEDWATCH SUBMITTED UNDER MANUFACTURE REPORT NUMBER 9617229-2024-0011675. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN/WILL BE REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THE REASON(S) FOR REOPERATION IS/ARE: "SAGGING" AND LATER DEFLATION. CLARIFICATION TO B3 PARTIAL DATE OF EVENT: 2017 WAS PROVIDED.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D9, H3, H6. DEVICE EVALUATION: BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINT CODES ARE: - DEFLATION: OBSERVED AN OPENING ASSESSED AS SURGICAL DAMAGE. AS PER THE INVESTIGATION PROCEDURE, CREASES AND WEAR ABRASION WERE OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 0

PATIENT CALLED TO REPORT LEFT SIDE "SAGGING" AND LATER DEFLATION. PHYSICIAN CONFIRMED EVENT. DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

PATIENT CALLED TO REPORT LEFT SIDE "SAGGING" AND LATER DEFLATION. PHYSICIAN CONFIRMED EVENT. DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
481088 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 2072081

Patients

Seq Age Sex Outcome Treatment
1 31 YR Female Required Intervention