FDA Adverse Event Death Summary report: N

NARKOMED

MDR report key: 2190488 · Received July 28, 2011

Report

Report Number
2510954-2011-00001
Event Type
Death
Date Received
July 28, 2011
Date of Event
June 28, 2011
Report Date
June 29, 2011
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC. (PNC)
Product Code
BSZ
PMA / PMN Number
K033498
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FACILITY WAS CONTACTED ON (B)(6) 2011 AND THE BIOMED TECHNICIAN SAID THAT HE PERFORMED A COMPLETE DAILY/PRE USE CHECKOUT OF THE DEVICE AFTER THE REPORTED INCIDENT AND THE DEVICE PASSED ALL TESTS. THE DEVICE WAS PLACED BACK INTO USE AND WAS ACTUALLY IN USE AT THE TIME OF THE PHONE CALL AND THERE WERE NO REPORTED PROBLEMS. THE DEVICE LOGS WERE DOWNLOADED AND WERE SENT TO THE MFR FOR ANALYSIS. THE ANALYSIS OF THE LOGS DID NOT INDICATE ANY DEVICE MALFUNCTIONING DURING THE PERIOD OF THE REPORTED INCIDENT. THE DEVICE LOGS DID INDICATE THAT THERE WERE VARIOUS PT PARAMETER ALARMS ANNUNCIATED DURING THE CASE. CONCLUSION: BASED ON THE INFO REPORTED BY THE BIOMED TECHNICIAN AND THE DEVICE LOG ANALYSIS, THERE IS NO INDICATION OF A DEVICE FAILURE/MALFUNCTION. THE DEVICE LOG ANALYSIS INDICATED THAT THE DEVICE OPERATED AS INTENDED DURING THE PERIOD OF THE REPORTED INCIDENT. ALARMS WERE APPROPRIATELY GENERATED WHEN MEASURED PT PARAMETERS EXCEEDED THE ALARM LIMITS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE THAT INVOLVED THE USE OF ROBOTICS, THE PT WENT INTO CARDIAC ARREST AND COULD NOT BE REVIVED. IN DISCUSSIONS WITH FACILITY PERSONNEL, IT WAS BELIEVED THAT DURING THE PROCEDURE, THE PT BEGAN TO BLEED INTERNALLY, POSSIBLY AS A RESULT OF THE ROBOTIC INSTRUMENT NICKING AN ARTERY. THE CLINICIAN ATTEMPTED TO PUMP MORE BLOOD INTO THE PT BUT THE PT COULD NOT BE REVIVED. THERE WAS NO ALLEGED DEVICE MALFUNCTION NOR WAS IT ALLEGED THAT THE DEVICE CAUSED OR CONTRIBUTED TO THE REPORTED ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NARKOMED ANESTHESIA MACHINE BSZ DRAEGER MEDICAL SYSTEMS, INC. (PNC) 6400 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Death