FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2190335 · Received July 28, 2011

Report

Report Number
3004209178-2011-82324
Event Type
Injury
Date Received
July 28, 2011
Date of Event
July 14, 2011
Report Date
July 14, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE OF 780MG/DL. IT WAS STATED THAT THE CUSTOMER WAS VOMITING, NAUSEA AND KETONES. IT WAS STATED THAT THE CUSTOMER WAS TREATED WITH MANUAL INJECTIONS. TROUBLESHOOTING WAS PERFORMED. THE PROGRAMMING ON THE INSULIN PUMP WAS CORRECT. RAN A FIXED PRIME TEST AND THE INSULIN EXITED. IT WAS STATED THAT THE CUSTOMER HAD SEVERAL CT SCANS AND MRI WHILE WEARING THE INSULIN PUMP. PERFORMED A HIGH PRESSURE AND DISPLACEMENT TEST AND THE TESTS PASSED. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722WWS

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization