PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2190268 · Received July 19, 2011

Report

Report Number
2032227-2011-01781
Event Type
Injury
Date Received
July 19, 2011
Date of Event
July 5, 2011
Report Date
July 7, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS, WITH A BLOOD GLUCOSE READING OF 501 MG/DL. IT WAS STATED THAT THE CUSTOMER WAS UNDER SOME STRESS, AND HAD A COLD THE WEEK BEFORE. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT THERE WAS NO TUBING CLAMP FOR THE HIGH PRESSURE TEST. A TUBING CLAMP WAS MAILED TO THE CUSTOMER, AND WAS ADVISED TO CALL BACK TO CONDUCT THE TEST. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-722LNAB

Patients

Seq Age Sex Outcome Treatment
1 13 YR Hospitalization UNOMEDICAL QUICK-SET PARADIGM INSULIN INFUSION SET| MMT-399, LOT 9201685