FDA Adverse Event Malfunction Summary report: N

APEX PUSH MONORAIL

MDR report key: 2190144 · Received August 4, 2011

Report

Report Number
2134265-2011-03216
Event Type
Malfunction
Date Received
August 4, 2011
Date of Event
July 15, 2011
Report Date
July 15, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: AS THE UNIT HAS NOT BEEN RETURNED, A TECHNICAL ANALYSIS COULD NOT BE PERFORMED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE A BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS CORONARY VESSEL. AN APEX PUSH MONORAIL 8X1.50MM BALLOON WAS ADVANCED TO THE TARGET VESSEL. DURING INFLATION THE BALLOON RUPTURED AT 8 ATM. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 APEX PUSH MONORAIL CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - GALWAY H7493896108150 13883169

Patients

Seq Age Sex Outcome Treatment
1