FDA Adverse Event Malfunction Summary report: N

ARCHITECT C4000 PROCESSING MODULE (INTEGRATED)

MDR report key: 21900757 · Received April 24, 2025

Report

Report Number
3016438761-2025-00246
Event Type
Malfunction
Date Received
April 24, 2025
Date of Event
April 8, 2025
Report Date
May 5, 2025
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
UDI-DI
00380740003753
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FIELD SERVICE REPRESENTATIVE (FSR) RESOLVED THE ISSUE BY REPLACING THE TUBING FROM THE ASPIRATION PUMP 14 TO THE T-FITTING. NO ADDITIONAL DISCREPANT RESULT ISSUES HAVE BEEN REPORTED SINCE THE PART WAS REPLACED. PART NUMBER 7-205558-02 TBG, ASP PUMP 14 TO T FTG (ROHS) WAS DETERMINED TO BE THE LIKELY CAUSE OF THE ISSUE. THE INSTRUMENT SERVICE HISTORY REVIEW REVEALED NO ADDITIONAL SERVICE TICKETS ASSOCIATED WITH DISCREPANT/ERRATIC ICT PATIENT RESULTS. THERE WERE NO ADDITIONAL SERVICE OR COMPLAINT ISSUES ON OR AROUND THE DATE THIS COMPLAINT WAS INITIATED THAT MAY HAVE CONTRIBUTED TO THIS ISSUE. A REVIEW OF TRACKING AND TRENDING OF THE CLINICAL CHEMISTRY SYSTEMS DID NOT IDENTIFY ANY SIMILAR ISSUES RELATED TO THE ARCHITECT C4000 PROCESSING MODULE, OR THE PART NUMBER 7-205558-02 TBG, ASP PUMP 14 TO T FTG (ROHS) OR DISCREPANT RESULTS AS DESCRIBED IN THIS COMPLAINT. THE DEVICE HISTORY RECORDS WERE REVIEWED, AND NO NON-CONFORMANCES OR POTENTIAL NON-CONFORMANCES WERE IDENTIFIED. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ARCHITECT C4000 PROCESSING MODULE, SERIAL NUMBER (B)(6), OR THE PART NUMBER 7-205558-02 TBG, ASP PUMP 14 TO T FTG (ROHS) WERE IDENTIFIED.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY DEPRESSED SODIUM RESULTS ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE PATIENT. THE SAMPLE WAS REPEATED WITH HIGHER RESULTS. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6), INITIAL RESULT = 115 MMOL/L REPEATED ON THE ISTAT = 138 REPEATED ON THE GEM = 138 NEW SAMPLE SAME PATIENT PROCESSED ON THE ARCHITECT = 133 MMOL/L. REFERENCE RANGE = 135-145 MMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY DEPRESSED SODIUM RESULTS ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE PATIENT. THE SAMPLE WAS REPEATED WITH HIGHER RESULTS. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) INITIAL RESULT = 115 MMOL/L REPEATED ON THE ISTAT = 138 REPEATED ON THE GEM = 138 NEW SAMPLE SAME PATIENT PROCESSED ON THE ARCHITECT = 133 MMOL/L. REFERENCE RANGE = 135-145 MMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.