FDA Adverse Event Malfunction Summary report: N

INRATIO2

MDR report key: 2190060 · Received June 29, 2011

Report

Report Number
2027969-2011-01435
Event Type
Malfunction
Date Received
June 29, 2011
Date of Event
June 2, 2011
Report Date
June 29, 2011
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
K072727
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: (B)(6) 2011, INRATIO METER: 3.40, REFERENCE: 2.50, MEAN: 2.95, CONFIDENCE LIMITS: 1.8 - 4.2. THE MEAN OF THE INRATIO METER AND COMPARATIVE SYS INR WERE CALCULATED. BOTH INRATIO AND REFERENCE VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO FURTHER INVESTIGATION IS REQUIRED. CONCLUSION: ANALYSIS OF THE CLIENT'S DATA FROM INRATIO AND REFERENCE TESTS REVEALED THAT TEST RESULT COMPARISON MET ACCURACY CRITERIA. CUSTOMER'S RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. (B)(4). DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (B)(4), NO FURTHER ACTION IS REQUIRED AT THIS TIME. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS USING THE INRATIO2 METER: RESULTS AS FOLLOWS: DATE: (B)(6) 2011, INRATIO: 3.4, LAB: 2.5. METER AND LAB TESTS WERE DONE 30 MINUTES APART.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO2 PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 100139 251118

Patients

Seq Age Sex Outcome Treatment
1