FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2190034
·
Received June 29, 2011
Report
- Report Number
- 3008642652-2011-00111
- Event Type
- Malfunction
- Date Received
- June 29, 2011
- Date of Event
- May 11, 2011
- Report Date
- June 28, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE OR WIRES EXPOSED) HAS BEEN CONFIRMED. UPON EVAL, THE TRUNK CABLE AND CONNECTOR WERE DAMAGED. THE DAMAGE TO THE TRUNK CABLE OCCURRED BETWEEN THE DISTRIBUTION NODE AND THE FRONT THERAPY ELECTRODE. THE CAUSE FOR THE DAMAGED CABLE AND CONNECTOR CANNOT BE POSITIVELY DETERMINED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CONNECTOR AND CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE DAMAGED HIS ELECTRODE BELT AND IT WOULD NO LONGER WORK. THE PT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |