FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2190014 · Received August 3, 2011

Report

Report Number
6000001-2011-16324
Event Type
Malfunction
Date Received
August 3, 2011
Date of Event
June 30, 2011
Report Date
July 28, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH FAILURE CODE F-66 WAS CONFIRMED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY A DEFECTIVE MAIN BATTERY. THE MAIN BATTERY WAS REPLACED TO CORRECT THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A FLOGARD INFUSION PUMP WITH FAILURE 66; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS CONDITION WAS FOUND IN THE ER UPON POWER UP. THERE WAS NO REPORTED PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1