ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-05591
- Event Type
- Malfunction
- Date Received
- August 3, 2011
- Report Date
- July 8, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. THE EVALUATION OF THE PRODUCT HAS NOT BEEN COMPLETED; THEREFORE, NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE BLACK BOX DOWNLOAD VERIFIED A BATTERY VOLTAGE DROP ON (B)(6) 2011. THE BATTERY WAS NOT RETURNED WITH THE PUMP FOR INVESTIGATION. THE BATTERY COMPARTMENT AND CAP WERE FOUND TO BE INTACT. THE PUMP POWERED ON NORMALLY AND WAS EXERCISED FOR 6 HOURS WITHOUT OVERHEATING OR ALARMS. THE PUMP ELECTRICAL CURRENTS WERE TESTED AND FOUND TO BE WITHIN SPECIFICATIONS. THE PUMP COVER WAS REMOVED TO INSPECT FOR INTERNAL MOISTURE AND NONE WAS FOUND. THE POWER FLEX, BATTERY CONNECTIONS AND INTERNAL COMPONENTS WERE TESTED AND NO DEFECTS WERE FOUND. NO PUMP OVERHEATING WAS DUPLICATED.
ON (B)(6) 2011, THE LAY-USER/PATIENT CONTACTED ANIMAS ALLEGING THAT THE PUMP FELT HOT TO TOUCH. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |