FDA Adverse Event
Malfunction
Summary report: N
DST
MDR report key: 218939
·
Received April 14, 1999
Report
- Report Number
- 9615727-1999-00001
- Event Type
- Malfunction
- Date Received
- April 14, 1999
- Date of Event
- February 4, 1999
- Report Date
- February 16, 1999
- Manufacturer
- SMV INTERNATIONAL
- Product Code
- IYX
- Removal / Correction Number
- UNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
GANTRY LEANING SLIGHTLY WITH HEAD ONE ON THE TOP AND HEAD TWO ON THE BOTTOM. INVESTIGATION MADE BY "F/S" DURING PREVENTIVE MAINTENANCE BY MEANS OF A LASER BEAM AND "COR" TESTS, AND A CLICKING SOUND FROM THE ROTATION LET THE "F/S" STOP THE SYSTEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DST | GAMMACAMERA | IYX | SMV INTERNATIONAL | DST | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |