FDA Adverse Event Malfunction Summary report: N

PEN NDL 32G 4MM PRO 100 BOX 1200 US

MDR report key: 21890856 · Received April 23, 2025

Report

Report Number
3023359743-2025-00302
Event Type
Malfunction
Date Received
April 23, 2025
Report Date
April 23, 2025
Manufacturer
EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
Product Code
FMI
UDI-DI
00382903205509
PMA / PMN Number
K212015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS MEDWATCH SUBMISSION IS BOTH AN INITIAL AND SUPPLEMENTAL FILING. INVESTIGATION OF THE RESULTS CAN BE SEEN BELOW: INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED, AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED AT THIS TIME.

Description of Event or Problem · 0

CONSUMER REPORTED THAT THE NEEDLES ARE NOT ATTACHING TO THE INSULIN PEN. CONSUMER DOES NOT RE-USE. LOT #: 4121867 CATALOG #: 320550 DATE OF EVENT: UNKNOWN SAMPLES: DISCARDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
674411 PEN NDL 32G 4MM PRO 100 BOX 1200 US Needle, hypodermic, single lumen FMI EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND 320550 4121087 00382903205509

Patients

Seq Age Sex Outcome Treatment
1 NA Male