FDA Adverse Event Injury Summary report: N

STYLE 15 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 21887083 · Received April 22, 2025

Report

Report Number
9617229-2025-06651
Event Type
Injury
Date Received
April 22, 2025
Date of Event
March 28, 2025
Report Date
April 22, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS RUPTURE WAS RECEIVED ON MAY 06,2025, WITH LOT NUMBER 2152001. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINT CODES ARE: ¿ RUPTURE: OBSERVED AN BROKEN ASSESSED AS FOLD FLAW OPENING. AS PER THE INVESTIGATION PROCEDURE, NON PENETRATING NICK WERE OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RUPTURE. THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RUPTURE. THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
628214 STYLE 15 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2152001

Patients

Seq Age Sex Outcome Treatment
1 72 YR Female Required Intervention