FDA Adverse Event Malfunction Summary report: N

SYRINGE CATHETER TIP 50ML

MDR report key: 21885185 · Received April 22, 2025

Report

Report Number
1911916-2025-00299
Event Type
Malfunction
Date Received
April 22, 2025
Date of Event
December 13, 2024
Report Date
April 23, 2025
Manufacturer
BECTON DICKINSON
Product Code
KYZ
UDI-DI
30382903096207
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

PR (B)(4) FOLLOW UP. IT WAS REPORTED THERE ARE PUNCTURES IN THE PLASTIC PACKAGING. TO AID IN THE INVESTIGATION, ONE PHOTO WAS RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. THE PHOTO SHOWS A PACKAGING BLISTER BOTTOM WEB WITH ROUNDED MARKS FROM THE PACKAGING PROCESS; THEY ARE NOT HOLES. NO OTHER INFORMATION COULD BE OBTAINED FROM THE PHOTO. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309620, LOT 4172759. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED.

Description of Event or Problem · 0

MATERIAL #:309620. BATCH#:4172759. IT WAS REPORTED BY CUSTOMER THAT THEY FOUND A PUNCTURES IN PLASTIC PACKAGING. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. PLEASE FOLLOW UP WITH (B)(6) FOR THE PRODUCT CONCERN OUTLINED BELOW WITHIN 3 BUSINESS DAYS; AND ADVISE OF ANY SPECIAL SHIPPING INSTRUCTIONS FOR PICK UP, COURIER OR DISCARDING OF PRODUCT. PLEASE INVESTIGATE AND FOLLOW UP WITH A WRITTEN REPORT FROM YOUR QUALITY ASSURANCE AS TO YOUR FINDINGS AND RESOLUTIONS. PRODUCT CONCERN TICKET NUMBER: (B)(4). DATE OF INCIDENT: (B)(6) 2024. HOSPITAL: (B)(6). DEPARTMENT: (B)(6). PRODUCT: SYR LL TIP 50CC. VMID: (B)(4). LOT NUMBER: MULTIPLE LOTS. PRODUCT EXPIRY DATE: 31-MAY-2029. NUMBER OF DEFECTIVE PRODUCT(S): 1. IS SAMPLE AVAILABLE: NO. CONCERN DESCRIPTION: LOOKS LIKE SMALL PUNCTURES IN PLASTIC PACKAGING. THANK YOU IN ADVANCE FOR YOUR FOLLOW-UP ON THIS PRODUCT CONCERN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
674533 SYRINGE CATHETER TIP 50ML SYRINGE, IRRIGATING (NON DENTAL) KYZ BECTON DICKINSON 4172759 30382903096207

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown