FDA Adverse Event
Malfunction
Summary report: N
OXY2MASK O2/ETCO2 ADULT
MDR report key: 21881996
·
Received April 22, 2025
Report
- Report Number
- 21881996
- Event Type
- Malfunction
- Date Received
- April 22, 2025
- Date of Event
- April 4, 2025
- Report Date
- April 7, 2025
- Manufacturer
- SOUTHMEDIC, INC
- Product Code
- CCK
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- WI
- Reporter Occupation
- RISK MANAGER
- Health Professional
- N
Narratives
Description of Event or Problem · 0
PER PROTOCOL, WE ARE SUPPOSED TO MONITOR ETCO2 [END-TIDAL CARBON DIOXIDE] ON ALL OF OUR PATIENTS WITH OSA [OBSTRUCTIVE SLEEP APNEA] WHO RECEIVE MODERATE SEDATION. WE CURRENTLY USE THE O2/ETCO2 ADULT OXYMASK TO BOTH PROVIDE OXYGEN AND MEASURE ETCO2. REFERENCE NUMBER OM-2125-14SLM. MULTIPLE NURSES HAVE HAD DIFFICULTY MEASURING ACCURATE CO2 [CARBON DIOXIDE] LEVELS ON OUR SEDATED PATIENTS, WHICH WE BELIEVE IS DUE TO THE EXCESSIVE LENGTH OF THE CO2 TUBING. WE ARE ALL GETTING READINGS BETWEEN 7-11 MMHG, WHEN THEY SHOULD BE 35-45 MMHG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1352214 | OXY2MASK O2/ETCO2 ADULT | ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE | CCK | SOUTHMEDIC, INC | OM-2125-14SLM | W1005506 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |