FDA Adverse Event Malfunction Summary report: N

PATIENT INTERFACE NIM4CPB1 NIM 4.0

MDR report key: 21879040 · Received April 22, 2025

Report

Report Number
1045254-2025-01164
Event Type
Malfunction
Date Received
April 22, 2025
Date of Event
February 19, 2025
Report Date
April 22, 2025
Manufacturer
MEDTRONIC XOMED INC.
Product Code
ETN
UDI-DI
00763000395902
PMA / PMN Number
K200759
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS FOUND THAT THE CUSTOMER COMPLAINT COULD NOT BE CONFIRMED, NO FAULT WITH HARD WIRED. SOFTWARE WAS UPDATED AND REPLACED BAD FUSE. CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: NIM4CM01, SERIAL/LOT #: (B)(6), UDI#: (B)(4). H3: PRODUCT ANALYSIS FOUND THAT THE CUSTOMER COMPLAINT WAS CONFIRMED, PMB AND EIE BOARD FAILURE. SOFTWARE WAS UPDATED TO V1.7.5. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED DURING PRE-OP THAT REOCCURRING CONNECTION ISSUES WITH PI BOX WHEN WIRED (PI BOX DOES NOT RECOGNIZE CABLE). UNUSUAL NOISE AND ERRORS WHEN BOOTING UP, "SELF TEST FAILED", AND DURING USE, OUT OF MEASUREMENT RANGE - CALIBRATING, WHEN NOT USING CAUTERY. THERE WAS NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1338461 PATIENT INTERFACE NIM4CPB1 NIM 4.0 STIMULATOR, NERVE ETN MEDTRONIC XOMED INC. NIM4CPB1 00763000395902

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown "SEE H11...."