DAVINCI XI
Report
- Report Number
- 2955842-2025-16268
- Event Type
- Malfunction
- Date Received
- April 22, 2025
- Date of Event
- March 27, 2025
- Report Date
- March 27, 2025
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110898
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE FSE WAS ABLE TO REPRODUCE THE COMPLAINT AND REPLACED THE VIO INTEGRATED ELECTROSURGICAL GENERATOR UNIT (IESU). THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE THE IESU TO PERFORM FAILURE ANALYSIS (FA). FA WAS ABLE TO CONFIRM REPORTED COMPLAINT VIA SYSTEM LOGS AND REPRODUCE THE ISSUE DURING IN-HOUSE TESTING. DURING VISUAL INSPECTION, FA FOUND THE UNIT HAD NO SIGNIFICANT SCRATCHES OR DENTS. THE FRONT BEZEL WAS IN DECENT CONDITION. THE UNIT WAS PLACED ON GOLDEN SYSTEM AND WAS RUN IN NORMAL MODE. C-34 ERROR OCCURRED ON START-UP AND DURING MONO COAG. FAILURE ANALYSIS CONCLUDED THAT NO ROOT CAUSE COULD BE ATTRIBUTED TO THIS ISSUE.
THE PROBABLE CAUSE OF ERROR C-34 IS ATTRIBUTED TO A FAULTY ELECTRICAL COMPONENT INSIDE THE ERBE GENERATOR. THIS ERROR INDICATES A LOWER VOLTAGE THAN EXPECTED DURING ENERGY ACTIVATION. INTUITIVE SURGICAL, INC. (ISI) CONTACTED THE MEDICAL ENGINEER AND OBTAINED THE FOLLOWING INFORMATION: THE SYSTEM FUNCTIONALITY WAS CHECKED UPON POWERING ON THE SYSTEM, AND IT POWERED ON WITHOUT ERRORS. THERE WAS NO ERROR WHEN THE GENERATOR WAS BOOTED UP IN THE MORNING, BUT WHEN STARTING TO USE, THERE WAS AN ERROR. THE CUSTOMER EXCHANGED THE GENERATOR FOR ANOTHER ONE TO COMPLETE THE PROCEDURE. PATIENT DEMOGRAPHIC INFORMATION WAS REQUESTED BUT NOT PROVIDED.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED TOTAL GASTRECTOMY SURGICAL PROCEDURE, THE ERBE GENERATOR HAD ERROR C-00 THAT OFTEN OCCURRED AND REBOOTING THE SYSTEM AND THE GENERATOR DID NOT RESOLVE THE ISSUE. INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) CHECKED THE LOG AND FOUND ERROR C-34 LOGGED MANY TIMES. AT THIS TIME, IT IS UNKNOWN IF THE CUSTOMER USED A THIRD-PARTY GENERATOR OR CONVERTED TO ANOTHER DA VINCI SYSTEM TO COMPLETE THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 911611 | DAVINCI XI | VISION SIDE SYSTEM | NAY | INTUITIVE SURGICAL, INC | 381121-30 | N/A | 00886874110898 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES. |