FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 21878992 · Received April 22, 2025

Report

Report Number
2955842-2025-16268
Event Type
Malfunction
Date Received
April 22, 2025
Date of Event
March 27, 2025
Report Date
March 27, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE FSE WAS ABLE TO REPRODUCE THE COMPLAINT AND REPLACED THE VIO INTEGRATED ELECTROSURGICAL GENERATOR UNIT (IESU). THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE THE IESU TO PERFORM FAILURE ANALYSIS (FA). FA WAS ABLE TO CONFIRM REPORTED COMPLAINT VIA SYSTEM LOGS AND REPRODUCE THE ISSUE DURING IN-HOUSE TESTING. DURING VISUAL INSPECTION, FA FOUND THE UNIT HAD NO SIGNIFICANT SCRATCHES OR DENTS. THE FRONT BEZEL WAS IN DECENT CONDITION. THE UNIT WAS PLACED ON GOLDEN SYSTEM AND WAS RUN IN NORMAL MODE. C-34 ERROR OCCURRED ON START-UP AND DURING MONO COAG. FAILURE ANALYSIS CONCLUDED THAT NO ROOT CAUSE COULD BE ATTRIBUTED TO THIS ISSUE.

Additional Manufacturer Narrative · 0

THE PROBABLE CAUSE OF ERROR C-34 IS ATTRIBUTED TO A FAULTY ELECTRICAL COMPONENT INSIDE THE ERBE GENERATOR. THIS ERROR INDICATES A LOWER VOLTAGE THAN EXPECTED DURING ENERGY ACTIVATION. INTUITIVE SURGICAL, INC. (ISI) CONTACTED THE MEDICAL ENGINEER AND OBTAINED THE FOLLOWING INFORMATION: THE SYSTEM FUNCTIONALITY WAS CHECKED UPON POWERING ON THE SYSTEM, AND IT POWERED ON WITHOUT ERRORS. THERE WAS NO ERROR WHEN THE GENERATOR WAS BOOTED UP IN THE MORNING, BUT WHEN STARTING TO USE, THERE WAS AN ERROR. THE CUSTOMER EXCHANGED THE GENERATOR FOR ANOTHER ONE TO COMPLETE THE PROCEDURE. PATIENT DEMOGRAPHIC INFORMATION WAS REQUESTED BUT NOT PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED TOTAL GASTRECTOMY SURGICAL PROCEDURE, THE ERBE GENERATOR HAD ERROR C-00 THAT OFTEN OCCURRED AND REBOOTING THE SYSTEM AND THE GENERATOR DID NOT RESOLVE THE ISSUE. INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) CHECKED THE LOG AND FOUND ERROR C-34 LOGGED MANY TIMES. AT THIS TIME, IT IS UNKNOWN IF THE CUSTOMER USED A THIRD-PARTY GENERATOR OR CONVERTED TO ANOTHER DA VINCI SYSTEM TO COMPLETE THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
911611 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-30 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES.