FDA Adverse Event Malfunction Summary report: N

ALINITY I TOTAL PSA REAGENT KIT

MDR report key: 21876468 · Received April 21, 2025

Report

Report Number
3008344661-2025-00045
Event Type
Malfunction
Date Received
April 21, 2025
Date of Event
April 3, 2025
Report Date
June 20, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MTF
UDI-DI
00380740210540
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07P92 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 7P92-21 / 31, WITH 510K/PMA/BLA NUMBER P910007.

Additional Manufacturer Narrative · 0

INVESTIGATION INTO THE ISSUE HAS IDENTIFIED PERFORMANCE SHIFTS COULD OCCUR WITH SPECIFIC LOTS OF ALINITY I TOTAL PSA REAGENT KIT, LIST NUMBER 07P92, INCLUDING LOT 73155FZ00. A FIELD ACTION HAS BEEN ISSUED AND THE IMPACTED LOTS ARE NOT DISTRIBUTED IN THE US. NO FURTHER INFORMATION WILL BE SUBMITTED. ADDITIONAL INFORMATION IN D10 CONCOMITANT PRODUCT.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I TOTAL PSA RESULTS. THE FOLLOWING RESULTS WERE PROVIDED: SPECIMEN (1) (B)(6) 2025 , SID (B)(6), (LOT 73155FZ00) INITIAL RESULT 5.397 NG/ML, REPEATED (LOT 72104FZ00) 3.989 NG/ML, NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I TOTAL PSA RESULTS. THE FOLLOWING RESULTS WERE PROVIDED: SPECIMEN (1) (B)(6) 2025 SID (B)(6) (LOT 73155FZ00) INITIAL RESULT 5.397 NG/ML, REPEATED (LOT 72104FZ00) 3.989 NG/ML, NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1161366 ALINITY I TOTAL PSA REAGENT KIT TOTAL, PROSTATE SPECIFIC AG (NONCOMPLEXED/COMPLEXED) FOR DET OF PROSTATE CANCER MTF ABBOTT IRELAND DIAGNOSTICS DIVISION 73155FZ00 00380740210540

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY I PROCESSING MODU, 03R65-01, (B)(6)