SC MIDWEST REG LXNGTN KY 1
Report
- Report Number
- 1718850-2011-00095
- Event Type
- Other
- Date Received
- July 22, 2011
- Date of Event
- June 21, 2011
- Report Date
- June 23, 2011
- Manufacturer
- SORIN GROUP USA, INC.
- Product Code
- DTZ
- PMA / PMN Number
- PREAMENDMENT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
THE FACILITY DID NOT PROVIDE THE LOT NUMBER, THEREFORE THE EXPIRATION DATE IS UNK. THE FACILITY DID NOT PROVIDE THE LOT NUMBER, THEREFORE THE MANUFACTURE DATE IS UNK. SORIN GROUP (B)(4) REC'D A REPORT THAT, TWO HRS INTO BYPASS, THE 1/2" BOOT SPLIT RESULTING IN A BLOOD LEAK. THE PIECE OF TUBING WAS REPLACED AND THE CASE CONTINUED. TWO MORE HRS INTO THE CASE, THE 1/2" TUBING SPLIT AGAIN CAUSING BLOOD LOSS. THE CASE WAS FINISHED W/O FURTHER ISSUES. THE TOTAL BLOOD LOSS FROM THE PT WAS ONE LITER AND THE PT WAS GIVEN ONE UNIT OF PACKED CELLS. THERE WAS NO REPORT OF PT INJURY. THE PRODUCT WAS RETURNED TO SORIN GROUP (B)(4) FOR EVAL. SORIN GROUP IS CONDUCTING AN INVESTIGATION. A F/U REPORT WILL BE FILED WHEN THE INVESTIGATION IS COMPLETE. SECTION 5, BLOOD CARDIOPLEGIA, IN THE PUMP INSTRUCTIONS FOR USE STATE "¿INCORRECT TUBING CLAMP INSERTS CAN RESULT IN SEVERE FAILURES DURING PERFUSION SINCE THE TUBING IS EITHER PINCHED OR NOT SEATED SECURELY IN THE PUMP¿.CHECK IF THE TUBINGS LIE EVENLY ALONG THE PUMP RACEWAY. THE TUBINGS SHOULD NEITHER BE TWISTED, CROSSED OR KINKED¿BOTH TUBINGS MUST LIKE EVENLY W/O KINKS OR TWISTS ALONG THE PUMP RACEWAY, MUST NOT BE KINKED OR CROSSED, MUST BE TIGHTLY SECURED IN THE TUBING CLAMP INSERTS AND MUST NOT HAVE ANY PLAY WITHIN THE PUMP¿.THE CORRECT ADJUSTMENT OF THE PUMP OCCLUSION IS AN ESSENTIAL PRECONDITION." THE SORIN GROUP HEART LUNG INSTRUCTIONS FOR USE INCLUDED IN THE HEART LUNG PACK STATE THAT FAILURE TO MAINTAIN PROPER OCCLUSION OF THE ROLLER PUMP CAN LEAD TO PREMATURE PUMP HEAD TUBING FAILURE, RESULTING IN LEAKS. THE IFU STATES TO USE APPROPRIATE TUBING INSERTS AND TO MAINTAIN THE NATURAL CURVATURE OF THE TUBING WHEN PLACING IT IN THE RACEWAY. FAILURE TO PROPERLY INSTALL THE PUMP TUBING MAY RESULT IN DAMAGING THE TUBING.
SORIN GROUP (B)(4) REC'D A REPORT THAT, TWO HRS INTO BYPASS, THE 1/2" BOOT SPLIT RESULTING IN A BLOOD LEAK. THE PIECE OF TUBING WAS REPLACED AND THE CASE CONTINUED. TWO MORE HRS INTO THE CASE, 1/2" TUBING SPLIT AGAIN CAUSING BLOOD LOSS. THE CASE WAS FINISHED W/O FURTHER ISSUES. THE TOTAL BLOOD LOSS FROM THE PT WAS ONE LITER AND THE PT WAS GIVEN ONE UNIT OF PACKED CELLS. THERE WERE NO REPORT OF PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SC MIDWEST REG LXNGTN KY 1 | CUSTOM PERFUSION PACK | DTZ | SORIN GROUP USA, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |