ARCHITECT ANTI-HBS REAGENT KIT
Report
- Report Number
- 3008344661-2025-00043
- Event Type
- Malfunction
- Date Received
- April 21, 2025
- Date of Event
- March 27, 2025
- Report Date
- May 27, 2025
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- LOM
- UDI-DI
- 00380740159931
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE COMPLAINT INVESTIGATION FOR A FALSELY ELEVATED RESULT INCLUDED A SEARCH FOR SIMILAR COMPLAINTS, AND THE REVIEW OF THE COMPLAINT TEXT, TRENDING DATA, IN-HOUSE TESTING, LABELLING, AND DEVICE HISTORY RECORDS. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. CLINICAL SPECIFICITY OF THE ARCHITECT ANTI-HBS ASSAY WAS EVALUATED WITH A RETAINED KIT OF LOT 65679FZ01 AND DEMONSTRATED ACCEPTABLE SPECIFICITY SPECIFICATIONS, CONFIRMING THAT THIS LOT IS MEETING THE ARCHITECT PRODUCT REQUIREMENT. TRENDING REVIEW DETERMINED NO TRENDS FOR THE ISSUE FOR THE PRODUCT. DEVICE HISTORY RECORD REVIEW ON LOT 65679FZ01 DID NOT SHOW ANY POTENTIAL NON-CONFORMANCES, OR DEVIATIONS. LABELLING AND LITERATURE WERE REVIEWED WHICH ADEQUATELY ADDRESSES THE ISSUE UNDER REVIEW. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ARCHITECT ANTI-HBS ASSAY, LOT NUMBER 65679FZ01WAS IDENTIFIED.
AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 7C18 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 1L82, WITH 510K/PMA/BLA NUMBER P050051.
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT ANTI-HBS RESULTS ON A 50-YR OLD MALE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: (B)(6) 2025 SID (B)(6) = 390.63 / 453.28 MIU/ML, ANOTHER PLATFORM IS NEGATIVE. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT ANTI-HBS RESULTS ON A 50-YR OLD MALE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: (B)(6) 2025 SID (B)(6) = 390.63 / 453.28 MIU/ML, ANOTHER PLATFORM IS NEGATIVE. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1570387 | ARCHITECT ANTI-HBS REAGENT KIT | TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM) | LOM | ABBOTT IRELAND DIAGNOSTICS DIVISION | 65679FZ01 | 00380740159931 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Male | ARC I2000SR INST, 03M74-02, (B)(6) | ARC I2000SR INST, 03M74-02, (B)(6) |