DAVINCI XI
Report
- Report Number
- 2955842-2025-16216
- Event Type
- Malfunction
- Date Received
- April 18, 2025
- Date of Event
- March 28, 2025
- Report Date
- March 28, 2025
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110898
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- 003
Narratives
INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE ERBE WAS ANALYZED AND FOUND TO HAVE THE ERROR C-34 ON ERBE. SYSTEM LOGS SHOWED ERRORS 25913, C-34, C-30, AND M-11. VISUAL INSPECTION REVEALED A CRACKED BEZEL AT THE TOP LEFT CORNER. C-34 ERRORS OCCURRED AT STARTUP, WHILE C-30 AND M-11 ERRORS APPEARED DURING MONO CUT MODE. THE UNIT WAS TESTED ON A GOLDEN SYSTEM IN NORMAL MODE. THE COMPLAINT WAS CONFIRMED BASED ON FAILURE ANALYSIS. THE PROBABLE CAUSE OF ERROR C-34 IS ATTRIBUTED TO A FAULTY ELECTRICAL COMPONENT INSIDE THE ERBE GENERATOR. THIS ERROR INDICATES A LOWER VOLTAGE THAN EXPECTED DURING ENERGY ACTIVATION.
AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (IESU) TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID NOT YET RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION. THE PROBABLE CAUSE OF ERROR C-34 IS ATTRIBUTED TO A FAULTY ELECTRICAL COMPONENT INSIDE THE ERBE GENERATOR. THIS ERROR INDICATES A LOWER VOLTAGE THAN EXPECTED DURING ENERGY ACTIVATION. THIS ERROR CAN BE RESOLVED THROUGH TROUBLESHOOTING OR REPLACEMENT OF THE GENERATOR.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE INTEGRATED ELECTROSURGICAL UNIT (IESU) ERBE SHOWED ERROR MESSAGE C-34. THE CUSTOMER WAS GUIDED TO CHECK IF FUSE AND VOLTAGE ARE SET ACCORDING TO THE LOCAL MAINS VOLTAGE. THE CUSTOMER WAS ADVISED TO SET AN EXTERNAL ENERGY DEVICE, BUT THE MEDICAL TEAM DECIDED TO CONVERT TO LAPAROSCOPIC SURGERY. THE SYSTEM COULD BE USED WITH AN EXTERNAL ENERGY DEVICE. THE PROCEDURE WAS CONVERTED TO LAPAROSCOPIC SURGERY WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE SITE AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE SYSTEM FUNCTIONALITY WAS CHECKED UPON POWERING ON THE SYSTEM AND THE SYSTEM INITIALLY POWERED ON WITHOUT ERRORS. THE CONVERSION DID NOT RESULT IN INCREASING PORT SIZE INCISION OR ADDING ADDITIONAL PORTS. THE PATIENT TOLERATED THE CONVERSION. THERE WAS NO INJURY TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30387 | DAVINCI XI | VISION SIDE SYSTEM | NAY | INTUITIVE SURGICAL, INC | 381121-24 | 00886874110898 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Female | DA VINCI INSTRUMENTS AND ACCESSORIES |