CD HORIZON SOLERA 5.5/6.0
Report
- Report Number
- 1030489-2025-01768
- Event Type
- Malfunction
- Date Received
- April 18, 2025
- Date of Event
- January 8, 2025
- Report Date
- April 18, 2025
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC
- Product Code
- HXX
- UDI-DI
- 00643169043640
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- 003
Narratives
H3. DEVICE EVALUATION SUMMARY: PRODUCT ANALYSIS #(B)(4): 5584111 LOT# K24G1499 VISUAL AND OPTICAL EXAMINATION IDENTIFIED IT APPEARS THAT PART OF THE TIP OF THE INSTRUMENT HAS BEEN BROKEN OFF. THE METAL DEFORMATION GIVES INDICATION THAT THE FAILURE WAS THE RESULT OF OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING THE DEVICES USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THE TIP OF THE DRIVERS WAS DAMAGED DURING NORMAL USE AND THE RATCHETING HANDLE NO LONGER WORKS. THERE WAS NO PATIENT INVOLVED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30158 | CD HORIZON SOLERA 5.5/6.0 | SCREWDRIVER | HXX | MEDTRONIC SOFAMOR DANEK USA, INC | 5584111 | K24G1499 | 00643169043640 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |