FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES

MDR report key: 21862937 · Received April 17, 2025

Report

Report Number
1917413-2025-00468
Event Type
Malfunction
Date Received
April 17, 2025
Date of Event
March 20, 2025
Report Date
June 13, 2025
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
GIM
UDI-DI
30382903678564
PMA / PMN Number
SEE H11
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY : BD RECEIVED TWO PHOTOS FOR INVESTIGATION. A VISUAL EXAMINATION OF THESE PHOTOS REVEALED BLACK PARTICLES ON THE INSIDE OF THE TUBE. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE: FOREIGN MATTER - NON-BIOLOGICAL. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE MEDICAL DEVICE TYPES REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL DEVICE TYPE IS AS FOLLOWS: D2B: MEDICAL DEVICE TYPE: JKA. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: BK230980, K213670, AND K231373. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING THE BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES, THERE WAS FOREIGN MATTER SEEN IN 92 TUBES. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING THE BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES, THERE WAS FOREIGN MATTER SEEN IN 92 TUBES. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
24647 BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE GIM BECTON, DICKINSON & CO. (BROKEN BOW) 4288089 30382903678564

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown