YPSOPUMP INSET
Report
- Report Number
- 3003442380-2025-06539
- Event Type
- Malfunction
- Date Received
- April 17, 2025
- Date of Event
- March 17, 2025
- Report Date
- May 5, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V
- Product Code
- FPA
- PMA / PMN Number
- K061374
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS H11: INVESTIGATION SUMMARY THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED. THE BATCH 6010631 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3 AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2 FOR THE INCORRECT INSERTION OF THE SOFT CANNULA. COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 3 ON REFERENCE SAMPLES, REFERENCE SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6007741 WAS MANUFACTURED ACCORDING TO THE WI VERSION 75 MANUFACTURED IN THE LINE 5, ON 01/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED; NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 06/MAY/2025 AGAINST MALFUNCTION CODE INCORRECT INSERTION OF THE SOFT CANNULA AND LOT 6010631 AND NO OTHER COMPLAINT HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN SPAIN. IT WAS REPORTED THAT PATIENT FACED HYPERGLYCEMIA DUE TO INCORRECT INSERTION OF THE INFUSION SE ON (B)(6) 2025. THE SITE OF INSERTION WAS ABDOMEN. THE PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1748131 | YPSOPUMP INSET | UNO INSET II 60/6 GREY Y-CAP 10PAK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V | 86-060-52B6 | 6010631 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |