FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 21854565 · Received April 17, 2025

Report

Report Number
2955842-2025-15539
Event Type
Malfunction
Date Received
April 17, 2025
Date of Event
March 23, 2025
Report Date
March 25, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874112298
PMA / PMN Number
K220023
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE INSTRUMENT WAS ANALYZED AND FOUND TO HAVE BLADE DAMAGE AT THE MIDDLE OF THE BLADE. ONE OF THE BLADE EDGES WERE INDENTED. THE CUTTING EDGE DID NOT HAVE CORROSION THAT WOULD HAVE CONTRIBUTED TO THE BLADE DAMAGE OR CUTTING FAILURE. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE OF NICKS AND GOUGES ON THE INSTRUMENT BLADES IS ATTRIBUTED TO USE-RELATED DAMAGE WHEN ATTEMPTING TO CUT INCOMPATIBLE MATERIAL (I.E., HARD OBJECTS SUCH AS BONE OR CARTILAGE) OR MISHANDLING THE INSTRUMENT. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE SITE COULD NOT CONTRIBUTE FURTHER TO ALL THIS BECAUSE THE ERROR WITH THE BROKEN SCISSORS WAS ONLY NOTICED IN THE CENTRAL STERILIZATION DEPARTMENT.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE MONOPOLAR CURVED SCISSORS INSTRUMENT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS (FA) TESTING AND THE FA IS STILL IN PROGRESS.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT AFTER A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE MONOPOLAR CURVED SCISSORS INSTRUMENT BROKE OUT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1682100 ENDOWRIST MONOPOLAR CURVED SCISSORS NAY INTUITIVE SURGICAL, INC 470179-21 K16241017 0214 00886874112298

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES