FDA Adverse Event Malfunction Summary report: N

ASPARTATE AMINOTRANSFERASE ACC. TO IFCC WITH PYRIDOXAL PHOSPHATE ACTIVATION

MDR report key: 21854485 · Received April 17, 2025

Report

Report Number
1823260-2025-01172
Event Type
Malfunction
Date Received
April 17, 2025
Date of Event
March 26, 2025
Report Date
May 15, 2025
Manufacturer
ROCHE DIAGNOSTICS
Product Code
CIT
UDI-DI
04015630920549
PMA / PMN Number
K951595
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COBAS 4000 C311 STAND ALONE SYSTEM SERIAL NUMBER IS (B)(6). A FIELD SERVICE ENGINEER (FSE) RAN PROBE WASHES AND VERIFIED THE GEAR PUMP PRESSURE WAS WITHIN RANGE. THE FSE CHECKED PROBES, WASHING, AND RINSE MECHANISM LEVELS. THEY RAN ROCHE QC'S AND BIORAD QC'S FOR THE AST. THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 0

SECTION D, DEVICE IDENTIFICATION WAS UPDATED. RELEVANT FIELDS OF SECTIONS D AND G WERE UPDATED. THE ALARM TRACE DID NOT CONTAIN ANY CONSPICUOUS EVENTS. THE ROCHE QC WAS PASSING BEFORE THE EVENT. THE BIORAD QC WAS WITHIN RANGE ON THE DAY OF THE EVENT BUT OUT OF RANGE ON (B)(6) 2025. THE INVESTIGATION WAS UNABLE TO DETERMINE A DIRECT ROOT CAUSE. HOWEVER, THE INSTRUMENT WAS VERIFIED, AND THE ISSUE APPEARS TO BE RESOLVED.

Description of Event or Problem · 0

THERE WAS AN ALLEGATION OF QUESTIONABLE ASTL/PYP RESULTS FROM THE COBAS 4000 C311 STAND ALONE SYSTEM. PATIENT 1'S INITIAL RESULT WAS 12.9 U/L, AND THE REPEAT RESULTS OF 20.3 U/L AND 21.7 U/L WITH A NEW CALIBRATION LOT ON 27-MAR-2025. PATIENT 2'S INITIAL RESULT WAS 16.2 U/L, AND THE REPEAT RESULTS OF 22.6 U/L AND 19.5 U/L WITH A NEW CALIBRATION LOT ON 27-MAR-2025. THE RESULT OF 22.6 U/L FOR PATIENT 2 WAS DEEMED TO BE CORRECT. THE CUSTOMER DISCOVERED THE QUESTIONABLE RESULTS DURING METHOD COMPARISON TESTING.