FDA Adverse Event
Malfunction
Summary report: N
ULTRASOUND
MDR report key: 21849848
·
Received April 16, 2025
Report
- Report Number
- 3009498591-2025-00028
- Event Type
- Malfunction
- Date Received
- April 16, 2025
- Date of Event
- November 5, 2024
- Report Date
- April 16, 2025
- Manufacturer
- SIEMENS MEDICAL SOLUTIONS USA, INC.
- Product Code
- IYN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
WHEN MEASURING D(E) -- > D (B) -- > THEN THE EI/B DISTANCE RATIO IS CALCULATING CORRECTLY.
Description of Event or Problem · 0
STRAIN IMAGING EI/B RATIO IS NOT CALCULATED ACCORDING TO MEASUREMENT LABELS. WORKFLOW FOR REPRODUCE: 18L6 BREAST EXAM, VT STRAINING IMAGING 1. MEASUREMENT MODE, ACTIVATE "EI/B RATIO - D" 2. MEASURE DISTANCE ON B MODE IMAGE -- > D (B) MEASUREMENT VALUE IS DISPLAYED ON IMAGE SCREEN. 3. MEASURE DISTANCE ON ELASTO IMAGE -- > D (E) MEASUREMENT VALUE IS DISPLAYED ON IMAGE SCREEN. 4. THE EI /B DISTANCE RATIO IS CALCULATED, BUT RESULT IS B/EI DISTANCE RATIO, NOT EI/B DISTANCE RATIO! (PLEASE REFER TO ATTACHED IMAGE).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 105747 | ULTRASOUND | IYN | SIEMENS MEDICAL SOLUTIONS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |