LIGAMAX-5MM
Report
- Report Number
- 3005075853-2011-03122
- Event Type
- Malfunction
- Date Received
- July 30, 2011
- Date of Event
- July 12, 2011
- Report Date
- July 15, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). JAWS. ADDITIONAL INFORMATION WAS REQUESTED. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT THE DEVICE WAS USED DURING A LAPAROSCOPIC CHOLECYSTECTOMY. THE DEVICE WAS FIRING THE CLIPS CROOKED AND WERE ODDLY FORMED. THREE OF THE CLIPS DROPPED INTO THE BELLY AND WERE RETRIEVED THROUGH THE TROCAR. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | H43U17 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |