FDA Adverse Event Malfunction Summary report: N

ENDOWRIST SP

MDR report key: 21846003 · Received April 16, 2025

Report

Report Number
2955842-2025-15270
Event Type
Malfunction
Date Received
April 16, 2025
Date of Event
March 24, 2025
Report Date
March 24, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
PMA / PMN Number
K182371
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE THE SINGLE PORT MONOPOLAR CAUTERY INSTRUMENT TO PERFORM FAILURE ANALYSIS (FA). FA WAS ABLE TO CONFIRM AND REPRODUCED THE REPORTED COMPLAINT. VISUAL INSPECTION WAS PERFORMED, AND THE INSTRUMENT WAS FOUND TO HAVE A CRACKED TUBE ADAPTER. NO MATERIAL APPEARS TO BE MISSING. ADDITIONAL OBSERVATION NOT REPORTED BY SITE: THE INSTRUMENT WAS FOUND TO HAVE A LODGED COMPONENT IN THE TUBE ADAPTER. THE LODGED COMPONENT APPEARS TO BE THE EXTRUDED TUBE FROM A TIP ACCESSORY. THIS FAILURE IS LIKELY DUE TO THE IMPROPER INSTALLMENT OR REMOVAL OF THE TIP ACCESSORY. THE PROBABLE ROOT CAUSE OF A DAMAGED TUBE ADAPTER IS ATTRIBUTED TO EITHER TIP ACCESSORY INSTALLATION ISSUES OR DAMAGE DURING USE. THIS ISSUE CAN BE RESOLVED BY USING AN ALTERNATE INSTRUMENT TO COMPLETE THE PROCEDURE.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) CONTACTED THE INITIAL REPORTER AND OBTAINED THE FOLLOWING INFORMATION: THE SINGLE PORT MONOPOLAR CAUTERY INSTRUMENT WAS INSPECTED PRIOR TO THE CASE STARTING. THE STAFF FOUND THAT THE INSTRUMENT TIP WAS DAMAGED. THE INSTRUMENT WAS NOT USED ON THE PATIENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED SURGICAL PROCEDURE THAT THE THREAD OF THE SINGLE PORT MONOPOLAR CAUTERY INSTRUMENT'S TIP WAS DAMAGED. THE PROCEDURE WAS COMPLETED WITH NO REPORTS OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2348947 ENDOWRIST SP MONOPOLAR CAUTERY INSTRUMENT NAY INTUITIVE SURGICAL, INC 430007-55 S10180802 0014

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES