FDA Adverse Event Injury Summary report: N

COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET

MDR report key: 21845985 · Received April 16, 2025

Report

Report Number
3002808486-2025-00074
Event Type
Injury
Date Received
April 16, 2025
Date of Event
March 19, 2025
Report Date
July 4, 2025
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURERS REF#: (B)(4). G4) SIMILAR TO DEVICE MARKETED UNDER 510(K)/PMA: K233680. INVESTIGATION IS STILL IN PROGRESS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

MANUFACTURERS REF# (B)(4). SUMMARY OF INVESTIGATIONAL FINDINGS: THE CELECT-PT FILTER WAS PLACED, BUT ONE OF THE FILTER LEGS PENETRATED CAVA. THE FILTER WAS NOT REMOVED. NO DEVICE WAS RETURNED FOR ANALYSIS AND NO IMAGING COULD BE OBTAINED. THEREFORE, BASED ON THE INFORMATION PROVIDED ONLY IT WOULD BE INAPPROPRIATE TO SPECULATE AT WHAT MAY OR MAY NOT HAVE CAUSED THE FILTER TO PENETRATE THE VENA CAVA DURING PLACEMENT. THE FILTER POSITION/ORIENTATION MAY HAVE BEEN SLIGHTLY AFFECTED DURING PLACEMENT, BUT ANY REASON WOULD BE PURE SPECULATION. THE COMPLAINT WILL BE REOPENED IF ADDITIONAL INFORMATION IS OBTAINED. THERE ARE ADEQUATE CONTROLS IN PLACE TO ENSURE THE DEVICE WAS MANUFACTURED TO SPECIFICATIONS. IT IS ASSESSED THAT BECAUSE NO NON-CONFORMANCES WERE DETECTED THERE IS EVIDENCE THAT THE DEVICE WAS MANUFACTURED IN COMPLIANCE WITH PROCEDURES AND ACCORDING TO SPECIFICATIONS. IN ADDITION, NO OTHER LOT RELATED COMPLAINTS HAVE BEEN RECEIVED FROM THE FIELD. IT IS THEREFORE CONCLUDED THAT THERE IS NO EVIDENCE THAT NON-CONFORMING PRODUCTS EXIST IN HOUSE OR IN FIELD. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THE FILTER IS IMPLANTED BUT IT IS EVIDENT THAT ONE OF THE LEGS PERFORATES THE CAVA. PATIENT OUTCOME: THE FILTER HAS NOT BEEN REMOVED YET.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2425529 COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE G34504 E4539151

Patients

Seq Age Sex Outcome Treatment
1 NA Female Required Intervention