FDA Adverse Event Malfunction Summary report: N

NEEDLE NS 30GA 7/8IN BSG W/O SIL W/O SHD

MDR report key: 21839174 · Received April 15, 2025

Report

Report Number
1911916-2025-00279
Event Type
Malfunction
Date Received
April 15, 2025
Date of Event
April 3, 2025
Report Date
April 17, 2025
Manufacturer
BECTON DICKINSON
Product Code
FMI
UDI-DI
50382903016643
PMA / PMN Number
UNKNOWN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(6). FOLLOW UP: IT WAS REPORTED THERE WAS A HARD PLASTIC BALL COVERING THE TIP OF THE CANNULA. AS A PHYSICAL SAMPLE WAS NOT RETURNED, A THOROUGH SAMPLE EVALUATION COULD NOT BE PERFORMED. TO AID IN THE INVESTIGATION, ONE PHOTO WAS PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. THE PHOTO SHOWS A METAL PART BENT INTO A CURVE. THE TIP IS ROUNDED AND THERE IS A DROPLET OF A WHITE SUBSTANCE. THE EPOXY CONDITION ON THE NEEDLE OCCURS IF THERE IS A MISFEEDING OF THE CANNULATOR. THE UNIT SHOWN IN THE PHOTO DOES NOT APPEAR TO BE AN ASSEMBLY MANUFACTURED BY THIS BD SITE OR HAS BEEN THROUGH A MANUFACTURING PROCESS BY THE CUSTOMER. THE PHYSICAL SAMPLE WOULD BE REQUIRED TO CONFIRM. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 301664, POSSIBLE LOT NUMBERS 3256516 AND 3151793. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THESE LOTS THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. BASED ON THE INVESTIGATION WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED.

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

MATERIAL # 301664 POTENTIAL BATCH # 3256516, 3151793 IT WAS REPORTED BY THE CUSTOMER THAT A HARD PLASTIC BALL WAS COVERING THE TIP OF THE CANNULA. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. - REPORTED ISSUE SUMMARIZED: IT WAS REPORTED THAT A HARD PLASTIC BALL WAS COVERING THE TIP OF THE CANNULA. - EVALUATION RESULTS: VISUAL INSPECTION WAS FOUND TO BE NON-CONFORMING FOR HAVING EPOXY ON THE TIP OF THE CANNULA. VENDOR: BECTON DICKINSON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
121959 NEEDLE NS 30GA 7/8IN BSG W/O SIL W/O SHD NEEDLE, HYPODERMIC, SINGLE LUMEN FMI BECTON DICKINSON UNKNOWN 50382903016643

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown