FDA Adverse Event Malfunction Summary report: N

COULTER LH 750 HEMATOLOGY ANALYZER

MDR report key: 2183630 · Received July 8, 2011

Report

Report Number
1061932-2011-00820
Event Type
Malfunction
Date Received
July 8, 2011
Date of Event
April 22, 2008
Report Date
April 25, 2008
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K061574
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS WITHIN QC SPECIFICATIONS WITH RESPECT TO CONTROLS AND REPRODUCIBILITY. CONTROLS WERE RUN ONCE PER SHIFT. CONTROLS WERE RUN BEFORE AND AFTER THIS INCIDENT. ON (B)(4) 2008, THE FIELD SERVICE ENGINEER EVALUATED THE ANALYZER AND VERIFIED THE INSTRUMENTS PERFORMANCE. RAW DATA ANALYSIS WAS CONDUCTED. ROOT CAUSE BASED ON RAW DATA ANALYSIS IS THAT THE SAMPLE SHOWS A LYMPHOCYTE DOMINANT PATTERN WITH MUCH SMALLER. NEUTROPHIL AND MONOCYTE POPULATIONS. BLAST FLAGS ARE BASED ON THE ABNORMAL PATTERNS OF NEUTROPHIL OR MONOCYTE POPULATIONS. THE ALGORITHM ALSO CHECKED FOR ABNORMALITY OF THE LYMPHOCYTE POPULATION AND GENERATED A VARIANT LYMPH FLAG FOR THE FIRST RERUN OF THE PATIENT SAMPLE. THE STATISTICS OF THE LYMPHOCYTE POPULATION WERE WITHIN NORMAL RANGE. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW CONDUCTED BETWEEN (B)(4) 2008 AND (B)(4), 2010 OF COMPLAINTS FOR ADDITIONAL REPORTABLE EVENTS.

Description of Event or Problem · 1

CUSTOMER REPORTED ERRONEOUS DIFFERENTIAL RESULTS WERE OBTAINED WHEN USING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE TEST RESULTS WERE NOT ACCOMPANIED BY INSTRUMENT GENERATED FLAGS. THE TEST RESULTS WERE DETERMINED TO BE ERRONEOUS BASED ON COMPARISON TO A MANUAL BLOOD SMEAR DIFFERENTIAL, WHICH CONTAINED BLAST CELLS. ERRONEOUS TEST RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. NO REPORTS OF DEATH OR SERIOUS INJURY, AND NO AFFECT TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER LH 750 HEMATOLOGY ANALYZER GKZ BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK