UNICEL DXI 800 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-02124
- Event Type
- Malfunction
- Date Received
- July 8, 2011
- Date of Event
- June 9, 2008
- Report Date
- June 9, 2008
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- MMI
- PMA / PMN Number
- K023764
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SVC ENGINEER (FSE) WAS ONSITE ON (B)(4) 2008. THE FSE PERFORMED PREVENTATIVE MAINTENANCE (PM) ON THE INSTRUMENT. THE FSE REPLACED THE SAMPLE PROBE BECAUSE IT APPEARED WORN OUT. THE FSE PERFORMED HIGH SENSITIVITY SYS CHECK, ROUTINE SYS CHECK, CARRYOVER AND A 50 REPLICATE PRECISION TEST WITH ACCUTNI LOW QUALITY CONTROL (QC) FOR VERIFICATION. ALL TESTS MET SET SPECS. THE FSE VERIFIED ALL REPAIRS PER ESTABLISHED PROCEDURES AND THE RESULTS MET PUBLISHED PERFORMANCE SPECS. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED FOR THIS EVENT. THIS IS ONE OF FIVE SEPARATE MDR REPORTS RELATED TO FIVE PT EVENTS ASSOCIATED WITH A SINGLE MALFUNCTION REPORT. REFERENCE MDR NUMBERS MDR 2122870-2011-02125, 02126, 02127, 02140 FOR ALL RELATED EVENTS. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS CONDUCTED BETWEEN (B)(4) 2008 THROUGH (B)(4) 2010 FOR ADDITIONAL REPORTABLE EVENTS.
THE CUSTOMER CONTACTED BECKMAN COULTER, INC (BCI) IN REGARDS TO ERRONEOUSLY ELEVATED ACCUTNI (TROPONIN I) RESULTS GENERATED ON A UNICEL DXI 800 ACCESS IMMUNOASSAY SYS FOR FIVE PTS. THE INITIAL ERRONEOUSLY ELEVATED RESULTS WERE REPORTED OUTSIDE THE LAB. THE CUSTOMER RE-RAN THE SAMPLES ON A DIFFERENT INSTRUMENT AND THE RESULTS WERE WITHIN NORMAL REFERENCE RANGE. IT IS NOT KNOWN IF THERE WAS ANY CHANGE TO PT TREATMENT. THERE IS NO INDICATION OF AN ADVERSE EVENT OR INDICATION OF ANY MEDICAL INTERVENTION TO PREVENT SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXI 800 IMMUNOASSAY SYSTEM | MMI | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | ACCESS ACCUTNI |