FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2183574 · Received July 11, 2011

Report

Report Number
1720753-2011-08889
Event Type
Malfunction
Date Received
July 11, 2011
Date of Event
June 28, 2011
Report Date
July 11, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REP CONDUCTED AN ONSITE INVESTIGATION. THE BATTERY PACK ASSEMBLY WAS REPLACED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A PRECHARGE FAILURE ERROR MESSAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1