FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 2183534 · Received July 11, 2011

Report

Report Number
9680959-2011-01554
Event Type
Malfunction
Date Received
July 11, 2011
Date of Event
June 27, 2011
Report Date
July 11, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE HIGH VOLTAGE CUBE NEEDED TO BE REPLACED. NO FURTHER INFORMATION IS AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A GENERATOR ERROR MESSAGE, AND THE SYSTEM WOULD NOT BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1