FDA Adverse Event Malfunction Summary report: N

PROXIMATE PPH PROCEDURE SET

MDR report key: 2183493 · Received July 29, 2011

Report

Report Number
3005075853-2011-03092
Event Type
Malfunction
Date Received
July 29, 2011
Date of Event
July 10, 2011
Report Date
July 10, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051301
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EG
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE WAS NOT RETURNED. THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, THERE WAS A PACKAGING LOT OR BATCH NUMBER PROVIDED BY THE USER FACILITY. WITH THIS INFORMATION, THE MANUFACTURING RELATED RECORDS WERE REVIEWED AND WE WERE UNABLE TO CONFIRM THE COMPLAINT NOR DETERMINE A MANUFACTURING OR DESIGN RELATED CAUSE FOR THE REPORTED EVENT. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING/BEFORE AN UNKNOWN PROCEDURE, PURSE STRING WAS DONE, INTRODUCTION OF STAPLER, FIRING, PART OF STAPLERS ONLY FIRED AND SMALL PART OF MUCOSA EXITED. (NOT CIRCUMFERENTIAL) ALTHOUGH PURSE COMPLETELY INSIDE STAPLER. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE PPH PROCEDURE SET STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA G4TN9J

Patients

Seq Age Sex Outcome Treatment
1