PROXIMATE PPH PROCEDURE SET
Report
- Report Number
- 3005075853-2011-03092
- Event Type
- Malfunction
- Date Received
- July 29, 2011
- Date of Event
- July 10, 2011
- Report Date
- July 10, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K051301
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EG
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). DEVICE WAS NOT RETURNED. THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, THERE WAS A PACKAGING LOT OR BATCH NUMBER PROVIDED BY THE USER FACILITY. WITH THIS INFORMATION, THE MANUFACTURING RELATED RECORDS WERE REVIEWED AND WE WERE UNABLE TO CONFIRM THE COMPLAINT NOR DETERMINE A MANUFACTURING OR DESIGN RELATED CAUSE FOR THE REPORTED EVENT. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING/BEFORE AN UNKNOWN PROCEDURE, PURSE STRING WAS DONE, INTRODUCTION OF STAPLER, FIRING, PART OF STAPLERS ONLY FIRED AND SMALL PART OF MUCOSA EXITED. (NOT CIRCUMFERENTIAL) ALTHOUGH PURSE COMPLETELY INSIDE STAPLER. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE PPH PROCEDURE SET | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | G4TN9J |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |