HANDPIECE, ELECTRIC DERMATOME
Report
- Report Number
- 0001526350-2025-00478
- Event Type
- Malfunction
- Date Received
- April 14, 2025
- Date of Event
- March 28, 2025
- Report Date
- September 30, 2025
- Manufacturer
- ZIMMER SURGICAL, INC.
- Product Code
- FZQ
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS COMPLAINT IS RECORDED BY ZIMMER BIOMET UNDER (B)(4). A FOLLOW UP/ FINAL REPORT WILL BE SUBMITTED ONCE INVESTIGATION IS COMPLETE. G2: FOREIGN COUNTRY - NETHERLANDS.
THIS COMPLAINT IS RECORDED BY ZIMMER BIOMET UNDER (B)(4). FOLLOWING SECTIONS WERE UPDATED OR CORRECTED: B4, B5, D4, D9, G3, G6, H2, H3, H4, H6 AND H11. REVIEW OF THE MOST RECENT REPAIR RECORD DETERMINED THE DEVICE WOULD NOT POWER ON; CABLE CONTACT ISSUE AND CORRODED MOTOR. THE MOTOR AND PLUG HARNESS ASSEMBLY WERE REPLACED AND RESOLVED THE REPORTED ISSUE. REVIEW OF THE DEVICE HISTORY RECORD IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. THE ROOT CAUSE OF THE REPORTED ISSUE IS ATTRIBUTED TO COMPONENT FAILURE AS THE DEVICE EXHIBITS WEAR AND TEAR FROM REPEATED USE. THE EVENT IS CONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR TRENDS.
IT WAS REPORTED THAT THE DERMATOME STARTS TO TURN SLOWER AND SLOWER. THE DERMATOME THEN BECOMES TOO HOT DURING PROLONGED USE. THE EVENT TIMING WAS OUTSIDE OF SURGERY. THERE WAS NO PATIENT INVOLVEMENT, AND NO HARM AND DELAY REPORTED. DUE DILIGENCE IS COMPLETE.
THERE WAS NO ADDITIONAL INFORMATION ASSOCIATED WITH THIS MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2269164 | HANDPIECE, ELECTRIC DERMATOME | DERMATOME | FZQ | ZIMMER SURGICAL, INC. | 62587639 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |