SYNCHRON® LX20 PRO ANALYZER
Report
- Report Number
- 2050012-2011-03322
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Date of Event
- June 29, 2011
- Report Date
- June 29, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K011213
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
A SERVICE REQUEST WAS GENERATED BY BEC CTS (CUSTOMER TECHNICAL SUPPORT) AND A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE (B)(6) 2011. FSE FOUNF A LEAK FROM THE HYDRO AND A NOISE FROM THE PUMP BOX. FSE REPLACED THE DAMPERS FOR VACUUM PUMP AND ALSO REPLACED SPLIT TUBING ON A VALVE. FSE PRIMED INSTRUMENT 50 TIMES AND NO FURTHER LEAKING WAS OBSERVED. THE BEC INTERNAL IDENTIFIER FOR THIS EVENT IS (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING A LEAK COMING FROM THE HYDROPNEUMATIC OF THE SYNCHRON LX20 PRO ANALYZER. THE CUSTOMER WAS UNABLE TO DETERMINE THE SOURCE OF IDENTITY OF THE FLUID BUT IT HAD THE VISCOSITY OF WASH CONCENTRATE. THE CUSTOMER ALSO STATED THAT THE AMOUNT WAS SMALL AND THEY COULD CONTAIN THE LEAK TO PREVENT SPILLAGE ONTO THE FLOOR. NO INJURY WAS REPORTED AND NO ERRONEOUS RESULTS WERE GENERATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® LX20 PRO ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | LX20 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |