FDA Adverse Event Malfunction Summary report: N

TINA-QUANT ANTISTREPTOLYSIN O

MDR report key: 21833077 · Received April 14, 2025

Report

Report Number
1823260-2025-01121
Event Type
Malfunction
Date Received
April 14, 2025
Date of Event
March 20, 2025
Report Date
August 11, 2025
Manufacturer
ROCHE DIAGNOSTICS
Product Code
GTQ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SERIAL NUMBER OF THE CUSTOMER'S COBAS 4000 C311 STAND ALONE SYSTEM IS (B)(6). THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 0

THE CUSTOMER RECEIVED A NEW QUESTIONABLE TINA-QUANT ANTISTREPTOLYSIN O ASSAY RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS 4000 C311 STAND ALONE SYSTEM. ON (B)(6) 2025: THE INITIAL RESULT FROM THE ANALYZER WAS 622 U/L. THE REPEAT RESULT FROM AN EXTERNAL LABORATORY WAS 455 U/L. THE FIELD SERVICE ENGINEER INSPECTED THE ANALYZER AND DID NOT FIND ANY ISSUES. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM BASED ON THE INFORMATION PROVIDED. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 0

ON THE FIELD SERVICE ENGINEER'S SERVICE VISIT, HE ALSO REPLACED THE SAMPLE PROBE AS A PROACTIVE MEASURE AND CALIBRATED THE WASH STATION.

Description of Event or Problem · 0

THE INITIAL REPORTER RECEIVED QUESTIONABLE TINA-QUANT ANTISTREPTOLYSIN O ASSAY RESULTS FROM ONE PATIENT SAMPLE TESTED ON THE COBAS 4000 C311 STAND ALONE SYSTEM. THE INITIAL RESULT FROM THE ANALYZER WAS 397 IU/ML, AND WAS CONFIRMED BY RERUNS. THE FIRST REPEAT RESULT FROM ANOTHER LABORATORY'S POLYMERASE CHAIN REACTION (PCR) ASSAY WAS 0.41 MG/L. THE SECOND REPEAT RESULT FROM ANOTHER LABORATORY'S ANALYZER WITH A SIMILAR LABORATORY METHOD TO THE REPORTED ASSAY WAS 276 U/L. THE INITIAL AND SECOND REPEAT RESULTS HAD THE SAME CLINICAL INTERPRETATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2349669 TINA-QUANT ANTISTREPTOLYSIN O ANTISTREPTOLYSIN - STREPTOLYSIN O REAGENT GTQ ROCHE DIAGNOSTICS 82332701

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown