PEN NDL 32G 4MM PRO 100 BOX 1200 US
Report
- Report Number
- 3023359743-2025-00283
- Event Type
- Malfunction
- Date Received
- April 14, 2025
- Report Date
- April 14, 2025
- Manufacturer
- EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
- Product Code
- FMI
- PMA / PMN Number
- K212015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- OTHER
Narratives
THIS MEDWATCH SUBMISSION IS BOTH AN INITIAL AND SUPPLEMENTAL FILING. INVESTIGATION OF THE RESULTS CAN BE SEEN BELOW: INVESTIGATION SUMMARY: SAMPLES WERE RECEIVED AND AN INVESTIGATION WAS PERFORMED. THIS IS THE 2ND COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AS BENT NON PATIENT END CANNULA AND BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED.
CONSUMER REPORTED PEN NEEDLE CLOG WHEN TAKING INJECTION. STATED, SHE PRIMES 1 UNIT ONLY EXPERIENCED THE PROBLEM WITH 5 PEN NEEDLES THIS MORNING BUT MORE WERE AFFECTED FROM THIS BOX BEFORE TODAY. CONSUMER DOES NOT KNOW THE DATE FOR OTHER PEN NEEDLES FAILING. LOT: 4051849. CATALOG: 320550. DATE OF EVENT: 2025-20-03. SAMPLES: YES. CONSUMER STATED, SHE WOULD PREFER THE ORIGINAL NANO PEN NEEDLES BUT THE PHARMACIST ONLY HAD THE 2ND GEN AVAILABLE. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1022629 | PEN NDL 32G 4MM PRO 100 BOX 1200 US | Needle, hypodermic, single lumen | FMI | EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND | 320550 | 4051849 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |