FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2183259 · Received July 28, 2011

Report

Report Number
6000001-2011-15377
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
June 30, 2011
Report Date
July 12, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ROOT CAUSE OF THE REPORTED CONDITION WAS ASSIGNED TO DEPLETED BATTERIES DUE TO USE/USER ERROR. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE HAS NOT BEEN SERVICED PRIOR TO THIS EVENT FOR THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH A MALFUNCTION OF DAMAGED BATTERY WAS CONFIRMED AND REPRODUCED. THE BAXTER PERSONNEL HAVE DETERMINED THAT THIS CONDITION IS DUE TO DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO RESOLVE THE REPORTED PROBLEM. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A MALFUNCTION OF DAMAGED BATTERY. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION REPORTED RELATED TO THE DEVICE. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1