FLEXCATH STEERABLE SHEATH
Report
- Report Number
- 3002648230-2011-00100
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Date of Event
- June 28, 2011
- Report Date
- June 29, 2011
- Manufacturer
- MEDTRONIC CRYOCATH LP
- Product Code
- DRA
- PMA / PMN Number
- K081049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED ISSUE WAS CONFIRMED THROUGH TESTING. DISSECTION SHOWED A LEAKING HEMOSTATIC VALVE. A FIELD ACTION WAS INITIATED IN JULY 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION. RESULTS OF EVALUATION OF RETURNED DEVICE WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
DURING A CRYOABLATION PROCEDURE PERFORMED IN (B)(6), THE ARCTIC FRONT CATHETER WAS INSERTED IN THE FLEXCATH STEERABLE SHEETH (CATHETER INTRODUCER) AND PLACED INTO THE LEFT ATRIUM. WHEN ASPIRATING THE SHEATH, THE PHYSICIAN REPORTED THAT THERE WERE AIR BUBBLES PRESENT. THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLEXCATH STEERABLE SHEATH | CATHETER, STEERABLE | DRA | MEDTRONIC CRYOCATH LP | 3FC12 | 26378 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00069 YR |