FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2183115 · Received July 28, 2011

Report

Report Number
2531779-2011-05407
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
March 24, 2011
Report Date
June 29, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

RECALL # 2531779-03/24/2010/003-R. THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED ON (B)(4) 2011. EVALUATION REVEALED A DAMAGED FORCE SENSOR SHIM SPOKE. THE PUMP WAS REWOUND, LOADED, PRIMED, AND EXERCISED WITH NO ALARMS OCCURRING. REVIEW OF THE PUMP HISTORY REVEALS LOSS OF PRIME WARNINGS ASSOCIATED WITH LOW FORCE. THE FORCE SENSOR CALIBRATION CHECK PASSED. THE CARTRIDGE WAS ALSO RETURNED TO ANIMAS. THE CARTRIDGE PASSES VISUAL INSPECTION AND A FORCE TEST WAS PERFORMED WITH NO FAILURES OCCURRING.

Description of Event or Problem · 1

EVALUATION REVEALED THAT THE FORCE SENSOR SPOKE SHIM WAS DAMAGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 7 YR