FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2183037
·
Received July 28, 2011
Report
- Report Number
- 2531779-2011-05406
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Report Date
- April 22, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(6). RECALL #: 2531779-03/24/2010/003-R. THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. THE PUMP WAS REWOUND, LOADED, PRIMED, AND EXERCISED WITH NO ALARMS OCCURRING. REVIEW OF THE PUMP HISTORY REVEALS TWO LOSS OF PRIME WARNINGS ASSOCIATED WITH ZERO FORCE AND TWO "NO CARTRIDGE DETECTED" WARNINGS. THE CARTRIDGE WAS ALSO RETURNED TO ANIMAS AND PASSES VISUAL INSPECTION. A FORCE TEST WAS PERFORMED ON THE CARTRIDGE WITH NO FAILURES OCCURRING.
Description of Event or Problem · 1
EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |