FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2183037 · Received July 28, 2011

Report

Report Number
2531779-2011-05406
Event Type
Malfunction
Date Received
July 28, 2011
Report Date
April 22, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). RECALL #: 2531779-03/24/2010/003-R. THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. THE PUMP WAS REWOUND, LOADED, PRIMED, AND EXERCISED WITH NO ALARMS OCCURRING. REVIEW OF THE PUMP HISTORY REVEALS TWO LOSS OF PRIME WARNINGS ASSOCIATED WITH ZERO FORCE AND TWO "NO CARTRIDGE DETECTED" WARNINGS. THE CARTRIDGE WAS ALSO RETURNED TO ANIMAS AND PASSES VISUAL INSPECTION. A FORCE TEST WAS PERFORMED ON THE CARTRIDGE WITH NO FAILURES OCCURRING.

Description of Event or Problem · 1

EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 19 YR