HOMECHOICE
Report
- Report Number
- 1423500-2011-09922
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Date of Event
- June 24, 2011
- Report Date
- July 5, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN, ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (RTB-CAPA-(B)(4)).
(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS IDENTIFIED IN THE PATIENT THERAPY LOG ON (B)(6) 2011 AT 03:39:45 WITH DRAIN VOLUME OF 3412 ML DURING CYCLE 3. THE FILL VOLUME IS 2000 ML. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED. THIS EVENT MEETS OVERFILL CRITERIA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |