HOMECHOICE
Report
- Report Number
- 1423500-2011-09912
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Date of Event
- June 20, 2011
- Report Date
- July 26, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE ASSIGNABLE CAUSE OF THE IIPV IDENTIFIED VIA DEVICE LOG REVIEW WAS DETERMINED TO BE INSUFFICIENT DRAIN, USE ERROR, TIDAL UF REMOVAL SET TOO LOW. A LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. NO ISSUES WERE IDENTIFIED DURING THE REVIEW OF THE ACTUAL DEVICE'S PREVIOUS SERVICE HISTORY RECORD (SHR) THAT MAY HAVE CONTRIBUTED TO THE REPORTED DIFFICULTY OR THE ADDITIONAL IIPV. THE DEVICE WAS DETERMINED TO MEET SPECIFICATIONS FOR THE IIPV FOUND IN THE LOGS. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE HAS BEEN ESCALATED TO CAPA.
(B)(4). THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS IDENTIFIED IN THE LOGS THAT OCCURRED ON DATE (B)(6) 2011 AT 06:53:14 WITH DRAIN VOLUME OF 3298 ML DURING CYCLE 6. THE FILL VOLUME IS 2000 ML. THIS EVENT MEETS OVERFILL CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |