FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2182995 · Received July 28, 2011

Report

Report Number
2531779-2011-05394
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
July 1, 2011
Report Date
July 2, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE BATTERY WAS NOT RETURNED WITH THE PUMP FOR INVESTIGATION. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED AND CORROSION WAS FOUND IN THE BATTERY COMPARTMENT. THE PUMP POWERED ON NORMALLY AND WAS EXERCISED FOR THREE HOURS WITH NO OVERHEATING OCCURRING. THE PUMP ELECTRICAL CURRENTS WERE TESTED AND FOUND TO BE WITHIN SPECIFICATIONS. THE PUMP WAS OPENED AND NO FURTHER EVIDENCE OF MOISTURE INGRESS WAS FOUND. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS FOUND TO BE FADED; THIS ISSUE DOES NOT EFFECT INSULIN DELIVERY FUNCTION.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT ON (B)(6) 2011, THE PUMP/BATTERY BECAME VERY WARM AFTER BEING EXPOSED TO SALT WATER. SHE STATED THAT THE SEAL ON THE BATTERY CAP WAS WEARING DOWN, AND SALT WATER GOT INTO THE BATTERY COMPARTMENT. SHE STATED, THERE WAS CORROSION ON THE BATTERY. THE PATIENT REPORTED THAT SHE DRIED OUT THE BATTERY COMPARTMENT AND PUT A NEW BATTERY IN. SHE STATED THAT SHE RESUMED THE PUMP WITHOUT ANY FURTHER ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 21 YR