FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60 ARTICULATING

MDR report key: 2182919 · Received July 28, 2011

Report

Report Number
3005075853-2011-03037
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
June 27, 2011
Report Date
June 30, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD LOADED ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE OPENED AND CLOSED WITHOUT ANY DIFFICULTIES NOTED AND NO CARTRIDGE RELOAD WAS RETURNED FOR ANALYSIS. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GASTRECTOMY PROCEDURE, THE DEVICE WOULD NOT STAPLE. THE STAPLING PROBLEM (NO MORE DETAILS) AND IT WAS IMPOSSIBLE TO UNLOCK THE DEVICE. THE SURGEON HAD TO UNLOCK IT MANUALLY. THE SURGEON USED ANOTHER LIKE DEVICE TO PERFORM THE PROCEDURE, WHICH ENDED WITHOUT FURTHER ISSUE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.