ECHELON*FLEX60 ARTICULATING ENDOSCOPIC LINEAR CUTTER, 60 MM
Report
- Report Number
- 3005075853-2011-03063
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Date of Event
- July 1, 2011
- Report Date
- July 6, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K081146
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS SHOWED THAT ONE EC60A DEVICE WAS RETURNED WITH NO VISUAL NON-CONFORMANCES AND LOADED WITH A 45 MM RELOAD. THE RELOAD WAS NOTED TO BE UNFIRED. PLEASE NOTE THAT A 60MM DEVICE IS DESIGNED TO WORK ONLY WITH 60MM CARTRIDGES, FOR FURTHER LOADING INSTRUCTIONS PLEASE REFER TO THE APEX 60MM IFU. IT SHOULD BE NOTED THAT WHEN ATTEMPTING TO FIRE A 60MM DEVICE WITH A 45MM CARTRIDGE RELOAD, THE DEVICE WILL LOCK OUT; IN ORDER TO OPEN A LOCKED DEVICE A REVERSE STROKE NEEDS TO BE PERFORMED TRIGGER TO TRIGGER IN ORDER TO RETURN THE KNIFE TO THE HOME POSITION (INDICATOR IN THE "0" POSITION) AND PRESS THE ANVIL RELEASE BUTTON TO OPEN. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A 60 MM RELOAD AND IT ACHIEVED A COMPLETE FIRE SEQUENCE WITHOUT ANY DIFFICULTIES NOTED. THE DEVICE FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE OPENED AND CLOSED AS INTENDED.
(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING A MINIMALLY INVASIVE ESOPHAGECTOMY PROCEDURE, THE DEVICE JAMMED ON TISSUE AND COULD NOT GET IT OFF. THERE ARE NO OTHER DETAILS AVAILABLE. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL FOLLOWUP: AFTER TALKING WITH THE SURGEON THE STAFF WAS USING AN EC60A AND LOADING A 45 CARTRIDGE AND THE DEVICE WOULD NOT FIRE THEN THE SURGEON WAS GIVEN ANOTHER EC60A AND CORRECT RELOAD AND WAS ABLE TO COMPLETE THE CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECHELON*FLEX60 ARTICULATING ENDOSCOPIC LINEAR CUTTER, 60 MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |